Why Standard Doors Are Not Enough

A standard wooden or steel door provides virtually zero radiation attenuation. In an X-ray or CT room, the door is the most vulnerable point in the shielding envelope. A properly specified lead lined door is therefore one of the most critical components of any radiological facility.

Under the Atomic Energy Authority of Sri Lanka (AEA) guidelines and IAEA Safety Reports Series No. 47, all access points to controlled radiation areas must provide shielding equivalent to the adjacent wall. Failure to meet this requirement is a common cause of AEA non-compliance during facility inspections.

Lead Equivalence: How Much Do You Need?

The required lead equivalence depends on the radiation type, energy, workload, and the occupancy of adjacent areas. As a general guide:

  • ◆Dental X-ray rooms: 1.0–1.5 mm Pb equivalent
  • ◆General diagnostic X-ray rooms: 1.5–2.0 mm Pb equivalent
  • ◆CT scanner rooms: 2.0–3.0 mm Pb equivalent
  • ◆Fluoroscopy suites: 2.0–2.5 mm Pb equivalent
  • ◆Nuclear medicine hot labs: 3.0–5.0 mm Pb equivalent
  • ◆Radiotherapy vaults (Linac): 10 mm Pb or concrete equivalent composite

Door Types and Construction

Lead lined doors are available in single leaf, double leaf, and sliding configurations. Single leaf doors are the most common for standard X-ray rooms. Double leaf doors are used where large equipment must pass through — for example CT scanners or radiotherapy equipment during installation.

Sliding doors are preferred in nuclear medicine and busy clinical environments where swing clearance is restricted. All door types require a continuous lead overlap at the frame perimeter to prevent radiation streaming at the edges — a detail often missed in cheaper installations.

NDT Inspectors fabricates all doors with 50 mm lead overlap at all edges and a continuous lead threshold seal. Frames are supplied as a matched set to ensure there are no gaps in the shielding envelope.

Lead Glass Viewing Windows

Most X-ray control areas require a lead glass observation window so the radiographer can observe the patient without entering the controlled area. Lead glass is available in lead equivalences from 1.5 mm to 20 mm Pb. The glass must be specified to the same lead equivalence as the door and wall.

NDT Inspectors supplies and installs matched door and window sets, ensuring the entire shielding system meets a single design specification verified by a qualified medical physicist.