Room Design 9 min read

CT Room Radiation Shielding Design: From Specification to Installation

End-to-end guide to CT room radiation shielding design — calculations, material selection, lead lined doors, lead glass windows and documented installation in Sri Lanka.

CT Shielding Design Parameters

CT scanner rooms require carefully calculated shielding based on the scanner's kVp range, mA settings, clinical workload, room geometry, and the occupancy of adjacent areas. Modern multidetector CT (MDCT) scanners operate at up to 140 kVp and can produce very high workloads — making accurate calculation essential.

Underspecification creates compliance risk and exposes staff to excess dose. Overspecification increases construction cost significantly. The goal is to reach the target dose constraint (typically 1 mSv/year for adjacent occupied areas) with the minimum required shielding.

Primary and Secondary Shielding

CT shielding is divided into primary (directly in the beam path) and secondary (scatter and leakage only) barriers. Primary barriers typically face the gantry and are the thickest.

A typical 64-slice CT installation might require: primary walls of 2.0–2.5 mmPb, secondary walls of 1.0–1.5 mmPb, ceiling of 0.5–1.0 mmPb (if occupied area above), and lead glass windows and lead lined doors at all penetrations.

Complete CT Room Package

NDT Inspectors provides end-to-end CT room shielding services — from initial design consultation and shielding calculations to material supply, fabrication, and documented installation with radiation verification.

  • ◆Primary wall lead lining or lead sheet installation (1.5–3.0 mmPb)
  • ◆Secondary wall and ceiling lead lining
  • ◆Lead lined viewing window with leaded glass (2mm–8mm Pb equivalent)
  • ◆Lead lined swing or sliding door with lead lined frame
  • ◆Lead lined service penetrations (electrical conduit, A/C ducting)
  • ◆Post-installation radiation survey with written compliance report