Why Charcoal Filtration Is Required in Iodine Therapy Rooms
Radioiodine (I-131) therapy involves administering large activities — typically 1–10 GBq — of iodine-131 to patients with thyroid cancer or hyperthyroidism. I-131 has a half-life of 8 days and decays with emission of beta particles and gamma rays. A significant fraction of the administered activity is excreted and can volatilise as radioiodine gas in the room air.
Without adequate filtration, radioiodine vapour exhausted from the treatment room contaminates building systems and the external environment, creating a public health and regulatory problem. The IAEA Safety Requirements GSR Part 3 and AEA Sri Lanka regulations require that exhaust air from iodine therapy rooms pass through a dedicated activated charcoal filter system before discharge.
How Activated Charcoal Filters Work
Activated charcoal (activated carbon) has an enormous internal surface area — typically 500–1500 m² per gram — created by steam activation of carbonaceous material. Radioiodine vapour adsorbs onto this surface with very high efficiency: a properly specified charcoal filter achieves iodine removal efficiencies of 99.9% or greater.
The filter is installed in the exhaust ductwork of the therapy room. Room air passes through the charcoal bed before being exhausted. After a period of use, the charcoal becomes saturated and must be replaced and disposed of as radioactive waste. The replacement interval depends on the room workload and the charcoal capacity.
System Design Parameters
Correct system design requires specifying the filter bed depth, face velocity, and charcoal grade. Key parameters include:
- ◆Face velocity: Typically 0.3–0.5 m/s. Higher velocities reduce contact time and efficiency.
- ◆Bed depth: Minimum 50 mm for standard iodine service; 100 mm recommended for high-workload rooms.
- ◆Charcoal grade: Impregnated activated carbon (potassium iodide or TEDA-impregnated) for radioiodine service.
- ◆Filter housing: Stainless steel with access panels for filter replacement and sampling ports for efficiency testing.
- ◆Bypass prevention: Positive-seal frame mounting to prevent air bypass at filter edges.
- ◆Monitoring: In-line iodine monitoring is recommended for high-workload facilities to detect filter breakthrough.
NDT Inspectors — Installed Systems
NDT Inspectors has designed, fabricated, and installed charcoal filter systems for iodine therapy rooms at multiple nuclear medicine facilities in Sri Lanka. All systems are supplied with filter efficiency test certificates and a recommended replacement schedule based on the facility's projected I-131 workload.